PNNL project 79550 provides for the analysis of neutron fluence monitors irradiated in the Advanced Test Reactor (ATR) at Idaho National Laboratory in accordance with MPO 00269673and Statement of Work No. 19704, Rev. 0, PNNL Analysis of NSUF Flux Capsules. This report is for the Boise State University (BSU) JW 15-8242 irradiations which were conducted in positions A6, A7 and A8 of the ATR. We previously reported data for this experiment after about 1 dpa in a report issued in January 2021 (1). This report is for longer irradiations to about 3 dpa. Other experiments included in the statement of work for this project will be reported separately. The neutron fluence monitors were prepared by PNNL and loaded into the BSU assemblies prior to irradiation. Following irradiation, the capsules were returned to PNNL for analysis. The neutron dosimetry capsules were opened, the flux wires were removed for gammaor x-ray analysis, and the measured activities were used to determine the activation rates for various activation products. Following suitable corrections, the measured activation rates were used to adjust calculated neutron spectra at each fluence monitor location. The adjusted neutron spectra were then used to determine displacement per atom (dpa) and gas production for irradiated materials.